Ad-Tech Medical Instrument Corporation Macro Micro Subdural Electrode

On 2013-03-13 the FDA classified a Class I recall of Ad-Tech Medical Instrument Corporation Macro Micro Subdural Electrode by Ad Tech Medical Instrument, citing the microelectrodes are not flush with the silastic substrate, may contain sharp edges capable of abrading the cortical surface, and may have remaining pieces which may fracture a….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0892-2013
ClassificationClass I
Statusterminated
Initiated2012-12-18
Published2013-03-13
Last updated2026-09-13 11:53 UTC
CompanyAd Tech Medical Instrument
DistributionUS

Product

Ad-Tech Medical Instrument Corporation Macro Micro Subdural Electrode. For temporary (<30 days) use with recording, monitoring and stimulation equipment, for the recording, monitoring and stimulation of electrical signals on the surface level of the brain. The recording of electrical activity supports definition of the location of epileptogenic foci and brain mapping.

Reason

The microelectrodes are not flush with the silastic substrate, may contain sharp edges capable of abrading the cortical surface, and may have remaining pieces which may fracture and remain in the cortical tissue with the electrode is explanted. Recall expanded on 5/2/2013 to include additional Macro Micro Subdural Electrodes.

Identifiers

code_info
REF numbers (catalog numbers): EG10A-SP10N-000, EG34B-SP10X-000, EGD7A-MP12N-000, EGE4A-SP10N-000, EGE4A-SP10X-000, Q…REF numbers (catalog numbers): EG10A-SP10N-000, EG34B-SP10X-000, EGD7A-MP12N-000, EGE4A-SP10N-000, EGE4A-SP10X-000, QG10B-SP10X-000, QG10C-SP10N-000, QG10D-SP10N-000, QG11A-SP10X-000, QG16A-SP10X-000, QG17A-SP10X-000, QG18A-SP10N-000, QG18A-SP10X-000, QG20B-SP10N-000, QG20B-SP10X-000, QG20C-SP10N-000, QG20C-SP10X-000, QG20D-SP10X-200, QG20E-SP10X-000, QG20F-SP10X-000, QG24A-SP10N-000, QG26A-SP10N-000, QG32A-SP10X-000, QG32B-SP10X-000, QG36A-SP10N-000, QG36A-SP10X-000, QG36B-SP10N-000, QG44A-SP10X-000, QG53A-SP10X-000, QG66A-SP85N-000, QG69A-SP10N-000, QG88A-SP10X-000, QG96A-SP10N-000, QG96A-SP10X-000, QG96C-SP10X-000, QG96D-SP10X-000, QGC6A-SP07N-000, QGC8A-SP10N-000, QGC8A-SP10X-000, QGC8B-SP10X-000, QGC8C-SP10X-000, QGC8D-SP10N-000, QGC8E-SP10X-000, QGC8F-SP10X-000, QS20A-SP10N-000, QS40B-SP10X-000, Lot numbers: 208140314, 208140343, 208140344, 208140353, 208140354, 208140362, 208140369, 208140374, 208140381, 208140382, 208140387, 208140390, 208140395, 208140398, 208140399, 20