FDA Device recall Z-0891-2022

ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of the ARCHITECT iSystem…

On 2022-04-13 the FDA classified a Class II recall of ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of the ARCHITECT iSystem… by Abbott Laboratories, citing this calibrator lot did not meet acceptance criteria during ongoing stability testing. When a calibration is performed with this lot, controls may exceed the specified range in th….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0891-2022
ClassificationClass II
Statusongoing
Initiated2022-03-08
Published2022-04-13
Last updated2026-09-13 11:56 UTC
CompanyAbbott Laboratories
DistributionUS

Product

ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of the ARCHITECT iSystem with STAT protocol capability when used for the quantitative determination of myoglobin in human serum and plasma.

Reason

This calibrator lot did not meet acceptance criteria during ongoing stability testing. When a calibration is performed with this lot, controls may exceed the specified range in the Instructions for Use (IFU).

Identifiers

code_infoUDI: (01)00380740003272(17)220531(10)166000

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.