FDA Device recall Z-0891-2022
ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of the ARCHITECT iSystem…
- FDA Device
- Class II
- ongoing
- Published 2022-04-13
On 2022-04-13 the FDA classified a Class II recall of ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of the ARCHITECT iSystem… by Abbott Laboratories, citing this calibrator lot did not meet acceptance criteria during ongoing stability testing. When a calibration is performed with this lot, controls may exceed the specified range in th….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0891-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-03-08 |
| Published | 2022-04-13 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Abbott Laboratories |
| Distribution | US |
Product
ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of the ARCHITECT iSystem with STAT protocol capability when used for the quantitative determination of myoglobin in human serum and plasma.
Reason
This calibrator lot did not meet acceptance criteria during ongoing stability testing. When a calibration is performed with this lot, controls may exceed the specified range in the Instructions for Use (IFU).
Identifiers
| code_info | UDI: (01)00380740003272(17)220531(10)166000 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Other recalls by Abbott Laboratories
- Alinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-20
FDA Device · Class II · 2025-12-24 · Abbott Laboratories - Brand Name: Alinity hq Analyzer Product Name: Alinity hq Analyzer Model/Catalog Number: 09P68-01 Software Version: 5.0 to 5.10.1 for OUS, 5…
FDA Device · Class II · 2025-11-19 · Abbott Laboratories - Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-13 Software Version: All software versions Product Descri…
FDA Device · Class II · 2025-04-30 · Abbott Laboratories - Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-01 Software Version: All software versions Product Descri…
FDA Device · Class II · 2025-04-30 · Abbott Laboratories - Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-10 Software Version: All software versions Product Descri…
FDA Device · Class II · 2025-04-30 · Abbott Laboratories
All Abbott Laboratories recalls
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0891-2022%22