FDA Device recall Z-0889-2015
ACCU-CHEK Tender II 13/60 10 + 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surfac…
- FDA Device
- Class II
- terminated
- Published 2015-01-07
On 2015-01-07 the FDA classified a Class II recall of ACCU-CHEK Tender II 13/60 10 + 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surfac… by Roche Diagnostics Operations, citing roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion se….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0889-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-11-20 |
| Published | 2015-01-07 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Roche Diagnostics Operations |
| Distribution | US |
Product
ACCU-CHEK Tender II 13/60 10 + 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections through the injection port. The indwelling catheter can also be securely attached to the infusion catheter by means of a proprietary click-lock connector for the infusion of drugs subcutaneously. The infusion set can be detached from the indwelling catheter, and the catheter capped to allow freedom from the infusion set and pump for showers, athletics or other activities.
Reason
Roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion set to become detached at the connect/disconnect location on the ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify
Identifiers
| code_info | Catalog: 04541464001 and lots: 5044364 and 5058502. |
|---|