RecallRadar

FDA Device recall Z-0888-2021

Philips Respironics V60 Ventilator Part Number 1053616 - Product Usage: intended to be used only with various combinations of Respironics-r…

On 2021-02-03 the FDA classified a Class II recall of Philips Respironics V60 Ventilator Part Number 1053616 - Product Usage: intended to be used only with various combinations of Respironics-r… by Respironics California, citing ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended peri….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0888-2021
ClassificationClass II
Statusterminated
Initiated2020-10-30
Published2021-02-03
Last updated2026-09-13 11:56 UTC
CompanyRespironics California
DistributionUS

Product

Philips Respironics V60 Ventilator Part Number 1053616 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.

Reason

Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC power.

Identifiers

code_info
Model: V60 Ventilator, Part Number 1053616 V60 Ventilator, English Opt: NONE Serial Numbers: 100087733 100087738…Model: V60 Ventilator, Part Number 1053616 V60 Ventilator, English Opt: NONE Serial Numbers: 100087733 100087738 100087932 100087999 100088155 100004031 100004071 100004073 100005244 100005350 100005351 100005599 100006209 100006210 100006217 100006406 100006701 100006702 100006881 100006883 100006886 100006888 100006891 100006929 100006933 100006939 100006940 100006941 100009417 100009501 100009512 100009513 100009549 100009550 100009553 100009554 100009556 100009566 100009679 100009760 100009761 100009762 100009768 100009769 100009770 100009777 100009812 100009816 100009854 100010206 100010242 100010390 100010514 100010515 100010517 100010519 100011308 100011343 100011344 100011468 100011740 100011741 100011744 100011749 100011750 100011751 100011754 100011791 100011794 100011799 100011801 100011820 100011821 100011822 100011823 100011824 100011825 100011826 100011827 100011828 100011829 100011830

Status history