FDA Device recall Z-0884-2013
Philips DigitalDiagnost X-Ray System with Eleva Software 2.1.3 Model numbers 712020, 712022, and 712082 Product Usage: Digital radiographic…
- FDA Device
- Class II
- terminated
- Published 2013-03-06
On 2013-03-06 the FDA classified a Class II recall of Philips DigitalDiagnost X-Ray System with Eleva Software 2.1.3 Model numbers 712020, 712022, and 712082 Product Usage: Digital radiographic… by Philips Healthcare, citing when the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0884-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-10-25 |
| Published | 2013-03-06 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Philips Healthcare |
| Distribution | US |
Product
Philips DigitalDiagnost X-Ray System with Eleva Software 2.1.3 Model numbers 712020, 712022, and 712082 Product Usage: Digital radiographic system
Reason
When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction. When the operator manually rotates the image back, electronic side markers (if programmed) are rotated also and as a result may be placed in a wrong position inside the image.
Identifiers
| code_info | Systems with Eleva SW version 2.1.3 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0884-2013%22