FDA Device recall Z-0882-2015
INRatio 2 Monitors - 0200431 Alere INRatio 2 PT/INR Professional Testing System - 0200432 Alere INRatio 2 PT/INR Home Monitoring System - 0…
- FDA Device
- Class I
- terminated
- Published 2015-01-14
On 2015-01-14 the FDA classified a Class I recall of INRatio 2 Monitors - 0200431 Alere INRatio 2 PT/INR Professional Testing System - 0200432 Alere INRatio 2 PT/INR Home Monitoring System - 0… by Alere San Diego, citing in certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). T….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0882-2015 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2014-12-05 |
| Published | 2015-01-14 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Alere San Diego |
| Distribution | US |
Product
INRatio 2 Monitors - 0200431 Alere INRatio 2 PT/INR Professional Testing System - 0200432 Alere INRatio 2 PT/INR Home Monitoring System - 0200433 Alere INRatio 2 PT/INR Home Monitoring System - 0200457 Alere INRatio 2 Replacement Monitor - 55112 Alere INRatio 2 Replacement Monitor - 55131 Alere INRatio 2 Replacement Monitor - 55128A Alere INRatio 2 PT/INR Professional Monitoring System The test systems contain individually packaged components needed to use the product, such as a monitor, user guide, quick reference guide training DVD or CD, Lancets (home monitoring kits contain an Autolet, Unilet, gauze pads, and alcohol wipes), power supply (home monitoring kits contain AA batteries), results log book and a warranty card. None of the components, except for the lancets, are sterile. Intended for in vitro diagnostic use for quantitative determination of international normalized ratio (INR) in fresh capillary whole blood to monitor the effect of warfarin on clotting time by health care professionals and home users.
Reason
In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can also occur if the instructions in the labeling for performing the test are not followed.
Identifiers
| code_info | All lots of part numbers 0200431, 0200432, 0200433, 0200457, 55112, 55131, and 55128A. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0882-2015%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.