LEGION AP Cutting Block Size 4, REF 71434409

On 2020-02-05 the FDA classified a Class II recall of LEGION AP Cutting Block Size 4, REF 71434409 by Smith And Nephew, citing multiple lots of LEGION AP Femoral Cutting Blocks were manufactured with an oversize green plunger hole and therefore do not meet their design specifications.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0878-2020
ClassificationClass II
Statusongoing
Initiated2019-12-16
Published2020-02-05
Last updated2026-09-13 11:55 UTC
CompanySmith And Nephew
DistributionDE, IA, NC, PA, TN

Product

LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 71434413. (Orthopaedic manual surgical instrument).

Reason

Multiple lots of LEGION AP Femoral Cutting Blocks were manufactured with an oversize green plunger hole and therefore do not meet their design specifications.

Identifiers

code_info
LEGION AP Cutting Block Size 4, Product Number 71434409, Lot Numbers: 17LM05399, 17LM07852, 17LM07852A; LEGION AP Cutt…LEGION AP Cutting Block Size 4, Product Number 71434409, Lot Numbers: 17LM05399, 17LM07852, 17LM07852A; LEGION AP Cutting Block Size 6, Product Number 71434411, Lot Numbers: 17LM05397, 17LM05398, 17LM07849; LEGION AP Cutting Block Size 7, Product Number 71434412, Lot Numbers: 17LM01746, 17LM05390, 17LM05394, 17LM07851; LEGION AP Cutting Block Size 8, Product Number 71434413, Lot Number: 17LM08442.