FDA Device recall Z-0877-2020
SIGNA Vibrant Nuclear Magnetic Resonance Imaging System, 1.5T Signa HDx, 1.5T Signa HDxt, Signa HDi - Product Usage: is a whole body magnet…
- FDA Device
- Class II
- terminated
- Published 2020-02-05
On 2020-02-05 the FDA classified a Class II recall of SIGNA Vibrant Nuclear Magnetic Resonance Imaging System, 1.5T Signa HDx, 1.5T Signa HDxt, Signa HDi - Product Usage: is a whole body magnet… by Ge Healthcare, citing it was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system s date and time setting are used to populate the DIC….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0877-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-12-23 |
| Published | 2020-02-05 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Ge Healthcare |
| Distribution | US |
Product
SIGNA Vibrant Nuclear Magnetic Resonance Imaging System, 1.5T Signa HDx, 1.5T Signa HDxt, Signa HDi - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
Reason
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system s date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.
Identifiers
| code_info | Mfg. Lot or Serial # System ID 22D-4-0868 901BMHCRITMR HDGOC1900010SC 209521SGMR HDGOC1900015SC 457771…Mfg. Lot or Serial # System ID 22D-4-0868 901BMHCRITMR HDGOC1900010SC 209521SGMR HDGOC1900015SC 4577711MR11 002776WHF 954916MR1 022D-4-1175 808243MR 857 LONGWOOD30MR2 00000290815MR0 508580BMR3 002801WHF 734246MR 002930WHF 845279MR 00000269040MR2 412749MR1 022D-4-1125 801RIVMR2 HDGOC1900003SC 206386SMR H05114 4815915MR11 00000277443MR8 304526SMR1 To be provided GON4796599 To be provided 424881MR01 To be provided 083027829345912 To be provided 083027862100408 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0877-2020%22