RecallRadar

FDA Device recall Z-0876-2024

SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-0510

On 2024-02-07 the FDA classified a Class II recall of SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-0510 by Howmedica Osteonics, citing product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material p….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0876-2024
ClassificationClass II
Statusongoing
Initiated2023-12-01
Published2024-02-07
Last updated2026-09-13 11:57 UTC
CompanyHowmedica Osteonics
DistributionUS

Product

SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-0510

Reason

Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

Identifiers

code_infoGTIN: 07613327034363 Lot Numbers: H43RDK, WR1RE3, 768NEY, 4J13P1

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0876-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.