RecallRadar

FDA Device recall Z-0875-2015

Trilogy¿ Fiber Metal Acetabular Shell Size 54 mm OD The Trilogy Acetabular System is indicated for either cemented or uncemented use in ske…

On 2014-12-31 the FDA classified a Class II recall of Trilogy¿ Fiber Metal Acetabular Shell Size 54 mm OD The Trilogy Acetabular System is indicated for either cemented or uncemented use in ske… by Zimmer Manufacturing B V, citing zimmer received a single complaint wherein a Trilogy® Fiber Metal Multi-Hole Acetabular Shell would not attach to the inserter/impactor instrumentation. Subsequent investigati….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0875-2015
ClassificationClass II
Statusterminated
Initiated2014-12-04
Published2014-12-31
Last updated2026-09-13 11:53 UTC
CompanyZimmer Manufacturing B V
DistributionUS

Product

Trilogy¿ Fiber Metal Acetabular Shell Size 54 mm OD The Trilogy Acetabular System is indicated for either cemented or uncemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.

Reason

Zimmer received a single complaint wherein a Trilogy® Fiber Metal Multi-Hole Acetabular Shell would not attach to the inserter/impactor instrumentation. Subsequent investigation found the polar-hole threads to be present but non-conforming to specifications, in that the minor diameter was undersized. Investigation also indicates that the issue is isolated and affects only lot 62714305, which

Identifiers

code_infoItem No. 00-6200-054-20; Lot 62714305

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0875-2015%22