FDA Device recall Z-0875-2015
Trilogy¿ Fiber Metal Acetabular Shell Size 54 mm OD The Trilogy Acetabular System is indicated for either cemented or uncemented use in ske…
- FDA Device
- Class II
- terminated
- Published 2014-12-31
On 2014-12-31 the FDA classified a Class II recall of Trilogy¿ Fiber Metal Acetabular Shell Size 54 mm OD The Trilogy Acetabular System is indicated for either cemented or uncemented use in ske… by Zimmer Manufacturing B V, citing zimmer received a single complaint wherein a Trilogy® Fiber Metal Multi-Hole Acetabular Shell would not attach to the inserter/impactor instrumentation. Subsequent investigati….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0875-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-12-04 |
| Published | 2014-12-31 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Zimmer Manufacturing B V |
| Distribution | US |
Product
Trilogy¿ Fiber Metal Acetabular Shell Size 54 mm OD The Trilogy Acetabular System is indicated for either cemented or uncemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.
Reason
Zimmer received a single complaint wherein a Trilogy® Fiber Metal Multi-Hole Acetabular Shell would not attach to the inserter/impactor instrumentation. Subsequent investigation found the polar-hole threads to be present but non-conforming to specifications, in that the minor diameter was undersized. Investigation also indicates that the issue is isolated and affects only lot 62714305, which
Identifiers
| code_info | Item No. 00-6200-054-20; Lot 62714305 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Other recalls by Zimmer Manufacturing B V
- VersaFx Femoral Fixation System Hex head compression screw 4.7 mm dia. 2 inch (51 mm) length 47116202400
FDA Device · Class II · 2016-05-25 · Zimmer Manufacturing B V - Mini MagnaFx Cannulated Screw Fixation System Small Cannulated Bone Screw, Fully Threaded, 5.0 mm Diameter, 20 mm Length 00114205020 Mini M…
FDA Device · Class II · 2016-05-25 · Zimmer Manufacturing B V - Hgp ii acetabular cup bone screw selftap 6.5x15mm 00662406515 Hgp ii acetabular cup bone screw selftap 6.5x20mm 00662406520 Hgp ii acetabul…
FDA Device · Class II · 2016-05-25 · Zimmer Manufacturing B V - Zimmer universal locking system (ULS) 3.5 mm cortical screw selftapping 18 mm length 47493501801
FDA Device · Class II · 2016-05-25 · Zimmer Manufacturing B V - Periarticular plating system, cancellous bone screw, partially threaded, hex head, 4.0 mm diameter, 16 mm length 47234702116
FDA Device · Class II · 2016-05-25 · Zimmer Manufacturing B V
All Zimmer Manufacturing B V recalls
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0875-2015%22