FDA Device recall Z-0873-2015
HemoCue¿ Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the…
- FDA Device
- Class II
- terminated
- Published 2014-12-31
On 2014-12-31 the FDA classified a Class II recall of HemoCue¿ Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the… by Radiometer America, citing batches of HemoCue Glucose 201 Microcuvettes show discoloration and provide results outside the specification at the end of their shelf life, approx. after 6 months. The HemoCue G….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0873-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-10-20 |
| Published | 2014-12-31 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Radiometer America |
| Distribution | US |
Product
HemoCue¿ Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the HemoCue Glucose 201 DM Analyzer (hereafter referred to as the HemoCue Glucose 201 Analyzer). HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials. Affected lots identified initially were 110705, 110706, 110715, 110716, 110717, 110718, 110719, and 110720. Subsequent investigation also identified as affected lots 1310227, 1310252, and 1310257. HemoCue¿ Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the HemoCue Glucose 201 DM Analyzer (hereafter referred to as the HemoCue Glucose 201 Analyzer). HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials.
Reason
Batches of HemoCue Glucose 201 Microcuvettes show discoloration and provide results outside the specification at the end of their shelf life, approx. after 6 months. The HemoCue Glucose 201 Analyzer with plasma conversion multiplies the measured whole blood glucose value by a factor of 1.11 and displays a plasma equivalent glucose result. HemoCue Glucose 201 Microcuvettes are for In Vitro Diag
Identifiers
| code_info | Model Number 110705: Lot numbers: 1310500, 1310501, 1310502, 1310503, 1310518, 1310519, 1310520, 1310521, 1310767, 13…Model Number 110705: Lot numbers: 1310500, 1310501, 1310502, 1310503, 1310518, 1310519, 1310520, 1310521, 1310767, 1310768, 1310769, 1310770, 1310771, 1310772, 1310779, 1310780, 1310781, 1310788, 1310789, 1310790, 1310799, 1311524, 1311525, 1311526, 1311527, 1311528, 1311529, 1311530, 1311531, 1311532, 1311533, 1311534, 1311535, 1311536, 1311537, 1311538, 1311539, 1311541, 1311542 Model Number 110706: Lot numbers: 1310207, 1310208, 1310209, 1310210, 1310211, 1310212, 1310213, 1310214, 1310215, 1310216, 1310228, 1310231, 1310232, 1310234, 1310235, 1310238, 1310239, 1310247, 1310249, 1310250, 1310256, 1310257, 1310258, 1310259, 1310260, 1311270, 1311271, 1311272, 1311273, 1311274, 1311279, 1311280, 1311281, 1311282, 1311283, 1311284, 1311291, 1311296, 1311297, 1311314, 1311315, 1311316, 1311317, 1311318, Model Number 110715: Lot numbers: 1310217, 1310218, 1310219, 1310221, 1310222, 1310229, 1310230, 1310237, 1310240, 1310241, 1310242, 1310251, 1310252, 1311267, 1311268, 1311 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0873-2015%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.