RecallRadar

FDA Device recall Z-0872-2024

SCORPIO RECESSED PATELLA-Intended for knee repalcment Part Number: 3044-0032

On 2024-02-07 the FDA classified a Class II recall of SCORPIO RECESSED PATELLA-Intended for knee repalcment Part Number: 3044-0032 by Howmedica Osteonics, citing product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material p….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0872-2024
ClassificationClass II
Statusongoing
Initiated2023-12-01
Published2024-02-07
Last updated2026-09-13 11:57 UTC
CompanyHowmedica Osteonics
DistributionUS

Product

SCORPIO RECESSED PATELLA-Intended for knee repalcment Part Number: 3044-0032

Reason

Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

Identifiers

code_infoGTIN: 07613327033823 Lot Numbers: 6LR4, R14V, N882

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0872-2024%22