RecallRadar

FDA Device recall Z-0864-2017

VITROS CRP Slides: 1) VITROS¿ Chemistry Products CRP Slides (250 slides per sales unit), Catalog # 1926740 and 2) , VITROS¿ Chemistry Produ…

On 2016-12-28 the FDA classified a Class III recall of VITROS CRP Slides: 1) VITROS¿ Chemistry Products CRP Slides (250 slides per sales unit), Catalog # 1926740 and 2) , VITROS¿ Chemistry Produ… by Ortho Clinical Diagnostics, citing CRP samples diluted using VITROS Chemistry Products Specialty Diluent, VITROS Chemistry Products FS Diluent Pack 3, or a sample with a low concentration of CRP had positively bias….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0864-2017
ClassificationClass III
Statusterminated
Initiated2016-10-26
Published2016-12-28
Last updated2026-09-13 11:54 UTC
CompanyOrtho Clinical Diagnostics
DistributionUS

Product

VITROS CRP Slides: 1) VITROS¿ Chemistry Products CRP Slides (250 slides per sales unit), Catalog # 1926740 and 2) , VITROS¿ Chemistry Products CRP Slides (90 slides per sales unit), Catalog # 1926740

Reason

CRP samples diluted using VITROS Chemistry Products Specialty Diluent, VITROS Chemistry Products FS Diluent Pack 3, or a sample with a low concentration of CRP had positively biased results if the endogenous level of CRP in the diluent is not subtracted from the final result. The IFUs do not provide specific instructions for how to account for endogenous CRP in the diluent when calculating the sample result. Orthos diluents used to dilute samples containing CRP are manufactured using human plasma based fluids that contain endogenous CRP. Without accounting for the CRP in the diluent and removing its contribution from the result, the reported result will include the CRP present in both the sample and diluent, possibly leading to positively biased results.

Identifiers

code_infoAffected Generations (GENs): 3, 11, 12, 30 through 36

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0864-2017%22