FDA Device recall Z-0859-2019
PageWriter TC30 w/ trolley Government Bundle, Product 860354, Software revisions up to and including A.07.05.22 to evaluate the electrocard…
- FDA Device
- Class II
- ongoing
- Published 2019-02-27
On 2019-02-27 the FDA classified a Class II recall of PageWriter TC30 w/ trolley Government Bundle, Product 860354, Software revisions up to and including A.07.05.22 to evaluate the electrocard… by Philips North America, citing philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0859-2019 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2019-01-09 |
| Published | 2019-02-27 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Philips North America |
| Distribution | US |
Product
PageWriter TC30 w/ trolley Government Bundle, Product 860354, Software revisions up to and including A.07.05.22 to evaluate the electrocardiogram of adult and pediatric patients
Reason
Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or nearby users.
Identifiers
| code_info | US41510645 US41510646 US71630792 USN1206382 USN1206383 USN1409879 USO1631023 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0859-2019%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.