RecallRadar

FDA Device recall Z-0858-2019

PageWriter TC70 w/o trolley Government Bundle, Product 860353, Software revisions up to and including A.07.05.22 to evaluate the electrocar…

On 2019-02-27 the FDA classified a Class II recall of PageWriter TC70 w/o trolley Government Bundle, Product 860353, Software revisions up to and including A.07.05.22 to evaluate the electrocar… by Philips North America, citing philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0858-2019
ClassificationClass II
Statusongoing
Initiated2019-01-09
Published2019-02-27
Last updated2026-09-13 11:55 UTC
CompanyPhilips North America
DistributionUS

Product

PageWriter TC70 w/o trolley Government Bundle, Product 860353, Software revisions up to and including A.07.05.22 to evaluate the electrocardiogram of adult and pediatric patients

Reason

Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or nearby users.

Identifiers

code_infoUS91412044 USO1513631

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0858-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.