FDA Device recall Z-0858-2019
PageWriter TC70 w/o trolley Government Bundle, Product 860353, Software revisions up to and including A.07.05.22 to evaluate the electrocar…
- FDA Device
- Class II
- ongoing
- Published 2019-02-27
On 2019-02-27 the FDA classified a Class II recall of PageWriter TC70 w/o trolley Government Bundle, Product 860353, Software revisions up to and including A.07.05.22 to evaluate the electrocar… by Philips North America, citing philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0858-2019 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2019-01-09 |
| Published | 2019-02-27 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Philips North America |
| Distribution | US |
Product
PageWriter TC70 w/o trolley Government Bundle, Product 860353, Software revisions up to and including A.07.05.22 to evaluate the electrocardiogram of adult and pediatric patients
Reason
Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or nearby users.
Identifiers
| code_info | US91412044 USO1513631 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0858-2019%22