FDA Device recall Z-0856-2021
Discovery NM 630 (Model Number H3101RH), NM 830 (Model Number H3910AC), 830 NM Cyber Defense (Model Number H3910AG) - Product Usage: intend…
- FDA Device
- Class II
- terminated
- Published 2021-01-20
On 2021-01-20 the FDA classified a Class II recall of Discovery NM 630 (Model Number H3101RH), NM 830 (Model Number H3910AC), 830 NM Cyber Defense (Model Number H3910AG) - Product Usage: intend… by Ge Healthcare, citing GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0856-2021 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-12-10 |
| Published | 2021-01-20 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Ge Healthcare |
| Distribution | n/a |
Product
Discovery NM 630 (Model Number H3101RH), NM 830 (Model Number H3910AC), 830 NM Cyber Defense (Model Number H3910AG) - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
Reason
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
Identifiers
| code_info | Discovery NM 630 (Model Number H3101RH), Serial numbers: 63GY50110, 63GY50119, 63GY50116, 63GZ50141, 63GY50120, 63GY…Discovery NM 630 (Model Number H3101RH), Serial numbers: 63GY50110, 63GY50119, 63GY50116, 63GZ50141, 63GY50120, 63GY50108, 63GY50121, 63GZ50142, 63GY50112, 63GZ50139, 63GY50107, 63GZ50138 NM 830 (Model Number H3910AC), Serial Numbers: 830Z60199, 830Z60182, 830Z60183, 830Z60192, 830Y60089, 830Y60105, 830Y60079, 830Z60207, 830Y60092, 830Y60102, 830Y60081, 830Z60200, 830Y60106, 830Z60201, 83EZ88001, 830Y60111, 830Y60094, 830Z60211, 830Y60101, 830Z60210, 830Y60118, 830Z60195, 830Z60194, 830Z60198, 830Z60181, 830Z60202, 830Y60097, 830Z60196, 830Z60180, 830Y60078, 830Y60115, 830Y60114, 830Y60108, 830Y60100, 830Y60087, 830Y60084, 830Z60188, 830Z60206, 830Y60112, 830Y60082, 830Y60093, 830Z60191, 830Y60107, 830Y60080, 830Z60190, 830Y60103, 830Z60193, 830Y60095, 830Y60119, 830Z60209, 830Z60205, 830Y60073, 830Y60120, 830Y60083, 830Y60116, 830Z60203, 830Y60085, 830Y60077, 830Z60184, 830Y60086, 830Z60187, 830Y60113, 830Z60208, 830Y60110, 830Z60204, 830Y60072, 830Z60179, 830Z60186 830 NM |
|---|