FDA Device recall Z-0856-2015

Lite DM (Lite Data Manager) Software Version(s): Versions 2.0.0.19 and lower Product Model #: D00A00224D The Lite DM (Lite Data Manager) is…

On 2014-12-31 the FDA classified a Class II recall of Lite DM (Lite Data Manager) Software Version(s): Versions 2.0.0.19 and lower Product Model #: D00A00224D The Lite DM (Lite Data Manager) is… by Horiba Instruments, citing HORIBA Medical is informing all HORIBA Medical Lite DM customers using software version 2.0.0.19 and lower of a software bug related to the potential for results received being ap….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0856-2015
ClassificationClass II
Statusterminated
Initiated2014-11-19
Published2014-12-31
Last updated2026-09-13 11:53 UTC
CompanyHoriba Instruments
DistributionUS

Product

Lite DM (Lite Data Manager) Software Version(s): Versions 2.0.0.19 and lower Product Model #: D00A00224D The Lite DM (Lite Data Manager) is a software application developed by HORIBA Medical intended to receive, store, and manage Startup, Patient, and Quality Control (QC) results from an ABX Micros 60 Hematology Analyzer or ABX Micros CRP 200 Hematology Analyzer. Only one instrument connection is allowed per each Lite DM workstation. The Lite DM application runs on a PC within a Windows Operating System platform. Data transmitted from the instrument to the Lite DM workstation is a Serial RS-232 unidirectional ABX Format communication. The Lite DM can act as a standalone data manager for data received from the instrument, or allow transmission of this data to a Host (LIS  Laboratory Information System or EMR  Electronic Medical Record system). Following the completion of each blood sample analysis performed on the connected HORIBA Medical hematology analyzer, results are then communicated (uni-directional) to the Lite DM workstation. Results received will display to the user in a specific color coding format indicating any normal or panic reference range violations. Results are then rejected or accepted by the user for final reporting - formatted to guidelines stipulated by CLIA. At no time are any results interpreted; only reported for review and appropriate action.

Reason

HORIBA Medical is informing all HORIBA Medical Lite DM customers using software version 2.0.0.19 and lower of a software bug related to the potential for results received being applied to the wrong order in the Lite DM Worklist. This bug has not been seen in software versions above V2.0.0.19 IMPACT: Potential for results received being applied to the wrong order in the Lite DM Worklist. The is

Identifiers

code_infoLot/Serial #(s): N/A Software Version(s): Versions 2.0.0.19 and lower. Expiration Date (s): N/A

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.