FDA Device recall Z-0851-2022
Hanger-Bar 2P 45cm
- FDA Device
- Class II
- ongoing
- Published 2022-04-06
On 2022-04-06 the FDA classified a Class II recall of Hanger-Bar 2P 45cm by Human Care Usa, citing the Hanger-Bar is being recalled due to the potential for the set screw loosening as a result of a manufacturing issue (failure to apply Loctite adhesive). As a result of the set….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0851-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-11-01 |
| Published | 2022-04-06 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Human Care Usa |
| Distribution | US |
Product
Hanger-Bar 2P 45cm
Reason
The Hanger-Bar is being recalled due to the potential for the set screw loosening as a result of a manufacturing issue (failure to apply Loctite adhesive). As a result of the set screw loosening, there is the potential for a fall hazard.
Identifiers
| upc | 881608800397 |
|---|---|
| code_info | Part Number 72754 UPC Code: 881608800397 EAN Code: 7331599999707 Lot Number: 33218501 |
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0851-2022%22