FDA Device recall Z-0851-2014
User Defined Method Flex Assignment/Siemens Dimension Vista 500 or Dimension Vista 1500 System with software version 3.5.1 or lower User De…
- FDA Device
- Class II
- terminated
- Published 2014-02-05
On 2014-02-05 the FDA classified a Class II recall of User Defined Method Flex Assignment/Siemens Dimension Vista 500 or Dimension Vista 1500 System with software version 3.5.1 or lower User De… by Siemens Healthcare Diagnostics, citing when utilizing the Routine Inventory screen to enter a User Defined Method (EMPTY) Flex the system may assign the User Defined Method Flex to a different Flex that is currently in….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0851-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-10-07 |
| Published | 2014-02-05 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | US |
Product
User Defined Method Flex Assignment/Siemens Dimension Vista 500 or Dimension Vista 1500 System with software version 3.5.1 or lower User Defined Method Flex Assignment/Siemens Dimension Vista 500 or Dimension Vista 1500 System is an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids.
Reason
When utilizing the Routine Inventory screen to enter a User Defined Method (EMPTY) Flex the system may assign the User Defined Method Flex to a different Flex that is currently in inventory on the system, and then use the incorrect Flex Cartridge to process the user defined method.
Identifiers
| code_info | Material numbers 10284473, 10488224, 10444801, and 10444802. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0851-2014%22