RecallRadar

FDA Device recall Z-0850-2018

Philips Network Firewall (Cisco ASA 5506), Model # 867098

On 2018-03-14 the FDA classified a Class II recall of Philips Network Firewall (Cisco ASA 5506), Model # 867098 by Philips Electronics North America, citing firewall installed with Philips IntelliVue Information Center iX or Information Center Classic may have a defective component, which may result in loss of connection to the Inform….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0850-2018
ClassificationClass II
Statusterminated
Initiated2017-10-05
Published2018-03-14
Last updated2026-09-13 11:54 UTC
CompanyPhilips Electronics North America
DistributionUS

Product

Philips Network Firewall (Cisco ASA 5506), Model # 867098

Reason

Firewall installed with Philips IntelliVue Information Center iX or Information Center Classic may have a defective component, which may result in loss of connection to the Information Center iX. The defect involves the clock signal component within the firewall. This component has a high probability of failing in appliances that have been running for greater than 18 months. If the clock signal component were to fail, the firewall will stop functioning, will not boot, and is not recoverable. This failure will result in loss of communication between devices that are separated by the firewall, which may cause the Information Center to reboot.

Identifiers

code_info
Serial Numbers for use in U.S.A.: JMX2030Y2B8 JMX2030Y2CR JMX2030Y2BN JMX2030Y2AY JMX2040Y2NZ JMX2022Y1U4 JMX204…Serial Numbers for use in U.S.A.: JMX2030Y2B8 JMX2030Y2CR JMX2030Y2BN JMX2030Y2AY JMX2040Y2NZ JMX2022Y1U4 JMX2040Y2NU JMX2045Y2L0 JMX2030Y2D9 JMX2030Y2B1 JMX2035Y0XC JMX2035Y0XT JMX2030Y2DY JMX2035Y0XA JMX2040Y2R2 JMX2031Y0EV JMX2039Y1FZ JMX2030Y2BT JMX2030Y2EE JMX2030Y2DM JMX194840WA JMX2014419Y JMX201441AB JMX201441A4 JMX2014Z155 JMX2014Z15C JMX2014Z15L JMX2039Y1H7 JMX2039Y1H5 JMX2039Y1HR JMX2039Y1JM JMX2022Y194 JMX2018Z009 JMX2022Y15F JMX2040Y2QP JMX2030Y2PZ JMX2045Y2E4 JMX20184033 JMX2030Y2A8 JMX2039Y1GB JMX2040Y2P9 JMX2030Y2E6 JMX2030Y2EG JMX2022Y181 JMX2022Y1A7 JMX2040Y2QW JMX2030Y2B6

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0850-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.