RecallRadar

FDA Device recall Z-0848-2025

8MM, POTTS SCISSORS, IS4000 REF 470001

On 2025-01-22 the FDA classified a Class II recall of 8MM, POTTS SCISSORS, IS4000 REF 470001 by Intuitive Surgical, citing due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0848-2025
ClassificationClass II
Statusongoing
Initiated2024-12-19
Published2025-01-22
Last updated2026-09-13 11:57 UTC
CompanyIntuitive Surgical
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY

Product

8MM, POTTS SCISSORS, IS4000 REF 470001

Reason

Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.

Identifiers

code_infoModel Number: 470001 Part Number: 470001-09 UDI-DI code: 00886874112120 Batch Numbers: K10230308 K11230308 K10230517

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0848-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.