RecallRadar

FDA Device recall Z-0844-2024

Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 88861919-51 SURGILON* 2-0 B…

On 2024-02-07 the FDA classified a Class II recall of Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 88861919-51 SURGILON* 2-0 B… by Covidien, citing specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0844-2024
ClassificationClass II
Statusongoing
Initiated2023-12-21
Published2024-02-07
Last updated2026-09-13 11:57 UTC
CompanyCovidien
DistributionUS

Product

Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 88861919-51 SURGILON* 2-0 BLK 7X75CMPCT 88861919-71 SURGILON* 1 BLK 7X75CM PCT

Reason

Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.

Identifiers

code_info
Model 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT UDI-DI 10884521071728, 20884521071725 Lots D1L1929RY D2B2243RY Model 888…Model 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT UDI-DI 10884521071728, 20884521071725 Lots D1L1929RY D2B2243RY Model 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT UDI-DI 10884521071742, 20884521071749 Lot D2B0176RY Model 88861919-51 SURGILON* 2-0 BLK 7X75CMPCT UDI-DI 10884521071766 Lots D2D2437RY UDI-DI 20884521071763 Lots D2D2437RY D2E0905RY D2E0916RY Model 88861919-71 SURGILON* 1 BLK 7X75CM PCT UDI-DI 20884521071794 Lots D1M0640RY D1M1494RY

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0844-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.