RecallRadar

FDA Device recall Z-0837-2021

Naturalyte Liquid Acid concentrate bottles, Product Code 08-3251-9 - Product Usage: used in conjunction with Fresenius Renal Technologies S…

On 2021-01-20 the FDA classified a Class II recall of Naturalyte Liquid Acid concentrate bottles, Product Code 08-3251-9 - Product Usage: used in conjunction with Fresenius Renal Technologies S… by Fresenius Medical Care Holdings, citing the deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously co….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0837-2021
ClassificationClass II
Statusongoing
Initiated2020-11-23
Published2021-01-20
Last updated2026-09-13 11:56 UTC
CompanyFresenius Medical Care Holdings
DistributionUS

Product

Naturalyte Liquid Acid concentrate bottles, Product Code 08-3251-9 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Reason

The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.

Identifiers

code_info
Lot Numbers: 20LXAC058 20LXAC078 20LXAC096 20NXAC011 20NXAC012 Additional Affected lots as of 3/17/21: 20LXAC060…Lot Numbers: 20LXAC058 20LXAC078 20LXAC096 20NXAC011 20NXAC012 Additional Affected lots as of 3/17/21: 20LXAC060 20NXAC026 20LXAC045 20LXAC043 20LXAC078 20NXAC012 20LXAC058 20LXAC096 20NXAC011 20LXAC059 20LXAC044 20LXAC079

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0837-2021%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.