FDA Device recall Z-0835-2014
DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee/Shoulder Blanket NS, REF T652NS, Rx only, Manufacturer DeRoyal Industries, Inc. 2…
- FDA Device
- Class II
- terminated
- Published 2014-02-05
On 2014-02-05 the FDA classified a Class II recall of DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee/Shoulder Blanket NS, REF T652NS, Rx only, Manufacturer DeRoyal Industries, Inc. 2… by Deroyal Industries, citing the unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0835-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-09-17 |
| Published | 2014-02-05 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Deroyal Industries |
| Distribution | US |
Product
DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee/Shoulder Blanket NS, REF T652NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.
Reason
The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.
Identifiers
| code_info | Lot Numbers: 32871821, 32976471, 33105247 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source