FDA Device recall Z-0834-2015

Fluorescence Imaging Procedure Kit, designed for use with Firefly Fluorescence Imaging for the da Vinci Si 3000 Surgical Systems

On 2014-12-31 the FDA classified a Class II recall of Fluorescence Imaging Procedure Kit, designed for use with Firefly Fluorescence Imaging for the da Vinci Si 3000 Surgical Systems by Intuitive Surgical, citing the storage conditions for indocyanine green (ICG) are not on the individual Fluorescence Imaging Procedure Kit box, the six-kit box, and the shipping box. If heat were to damage….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0834-2015
ClassificationClass II
Statusterminated
Initiated2014-11-21
Published2014-12-31
Last updated2026-09-13 11:53 UTC
CompanyIntuitive Surgical
DistributionUS

Product

Fluorescence Imaging Procedure Kit, designed for use with Firefly Fluorescence Imaging for the da Vinci Si 3000 Surgical Systems. The Intuitive Surgical Endoscopic Instrument Control System (Intuitive Surgical da Vinci Surgical System Model IS3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.

Reason

The storage conditions for indocyanine green (ICG) are not on the individual Fluorescence Imaging Procedure Kit box, the six-kit box, and the shipping box. If heat were to damage the material, the result would be a loss of fluorescence.

Identifiers

code_infoPN 950156-01

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0834-2015%22