RecallRadar

FDA Device recall Z-0834-2014

Aquarius (TM) TEMPERATURE THERAPY, REF T650, Rx Only, Manufactured for DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, Made in…

On 2014-02-05 the FDA classified a Class II recall of Aquarius (TM) TEMPERATURE THERAPY, REF T650, Rx Only, Manufactured for DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, Made in… by Deroyal Industries, citing the unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0834-2014
ClassificationClass II
Statusterminated
Initiated2013-09-17
Published2014-02-05
Last updated2026-09-13 11:53 UTC
CompanyDeroyal Industries
DistributionUS

Product

Aquarius (TM) TEMPERATURE THERAPY, REF T650, Rx Only, Manufactured for DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, Made in China. Physical Therapy.

Reason

The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.

Identifiers

code_info
Lot Numbers: 31480713, 31480721, 31968743, 31971619, 31990844, 32005711, 32018491, 32037764, 32118020, 32710832, 32736…Lot Numbers: 31480713, 31480721, 31968743, 31971619, 31990844, 32005711, 32018491, 32037764, 32118020, 32710832, 32736223, 33053264, 33105221, 33141176, 33164204, 33164407, 33276273, 33329911

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0834-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.