Naturalyte Liquid Acid concentrate bottles, Product Code 08-2201-5 - Product Usage: used in conjunction with Fresenius Renal Technologies S…

On 2021-01-20 the FDA classified a Class II recall of Naturalyte Liquid Acid concentrate bottles, Product Code 08-2201-5 - Product Usage: used in conjunction with Fresenius Renal Technologies S… by Fresenius Medical Care Holdings, citing the deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously co….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0832-2021
ClassificationClass II
Statusongoing
Initiated2020-11-23
Published2021-01-20
Last updated2026-09-13 11:56 UTC
CompanyFresenius Medical Care Holdings
DistributionUS

Product

Naturalyte Liquid Acid concentrate bottles, Product Code 08-2201-5 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Reason

The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.

Identifiers

code_infoLot Numbers: 20LXAC086 20LXAC099 20NXAC014 Additional Affected lots as of 3/17/21: 20LXAC061 20LXAC062 20LXAC099 20LXAC086 20NXAC014