FDA Device recall Z-0832-2013

CareFusion SmartSite Gravity Blood Set Model 42082E. Used to administer blood and blood products through a needle or catheter inserted into…

On 2013-02-27 the FDA classified a Class II recall of CareFusion SmartSite Gravity Blood Set Model 42082E. Used to administer blood and blood products through a needle or catheter inserted into… by Carefusion 303, citing careFusion is recalling the SmartSite¿ Gravity Blood Set as a result of a potential for separation at the connection of the drip chamber and tubing . The separation may be observe….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0832-2013
ClassificationClass II
Statusterminated
Initiated2012-12-20
Published2013-02-27
Last updated2026-09-13 11:53 UTC
CompanyCarefusion 303
DistributionUS

Product

CareFusion SmartSite Gravity Blood Set Model 42082E. Used to administer blood and blood products through a needle or catheter inserted into the patient's artery or vein.

Reason

CareFusion is recalling the SmartSite¿ Gravity Blood Set as a result of a potential for separation at the connection of the drip chamber and tubing . The separation may be observed at priming or observed during use. If a separation between the drip chamber and tubing occurs, this could cause a delay in treatment.

Identifiers

code_infoLot number 12055225

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.