FDA Device recall Z-0828-2024
Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LOWER PELVIC PACK, kit number AGLP52L
- FDA Device
- Class II
- ongoing
- Published 2024-02-07
On 2024-02-07 the FDA classified a Class II recall of Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LOWER PELVIC PACK, kit number AGLP52L by American Contract Systems, citing during an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting doc….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0828-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-11-29 |
| Published | 2024-02-07 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | American Contract Systems |
| Distribution | US |
Product
Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LOWER PELVIC PACK, kit number AGLP52L; b) OB/GYN DAVINCI PACK A RG BASIN, kit number HGOB98G; c) DAVINCI GYN PACK, kit number HHDV17; d) KIT, GYNE ROBOTIC, kit number MMGR44D; e) KIT, MAJOR LITHOTOMY, kit number MMJL16E; f) GYN PACK, kit number SMGY46; g) Lithotomy Pack, kit number THLH71C; h) ASC HYSTEROSCOPY PACK (PS 020634), kit number UIHY16AA; i) GYN LAPAROSCOPY HPES, kit number WELP67C; j) GYN LAPAROSCOPY HPES, kit number WELP67C-01; k) GYN LAPAROSCOPY HPES, kit number WELP67C-02
Reason
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
Identifiers
| code_info | a) AGLP52L, UDI/DI 00191072144119, Lot Numbers: 933231, exp. 3/8/2024; 942231, exp. 2/27/2024; 947231, exp. 2/22/2024;…a) AGLP52L, UDI/DI 00191072144119, Lot Numbers: 933231, exp. 3/8/2024; 942231, exp. 2/27/2024; 947231, exp. 2/22/2024; 966231, exp. 2/3/2024; 980231, exp. 1/20/2024; 681221, exp. 11/15/2023; 707221, exp. 10/20/2023; b) HGOB98G, UDI/DI none, Lot Numbers: 667221, exp. 11/29/2023; 977231, exp. 1/23/2024; 952231, exp. 1/31/2024; 896231, exp. 1/31/2024; c) HHDV17, UDI/DI 00191072170125, Lot Numbers: 766221, exp. 8/22/2023; 731221, exp. 9/26/2023; 682221, exp. 11/14/2023; 638221, exp. 12/28/2023; 946231, exp. 2/23/2024; d) MMGR44D, UDI/DI none, Lot Numbers: 709221, exp. 10/18/2023; 650221, exp. 12/16/2023; 891231, exp. 4/19/2024; e) MMJL16E, UDI/DI none, Lot Numbers: 989231, exp. 1/11/2024; 920231, exp. 3/21/2024; f) SMGY46, UDI/DI 00191072188458, Lot Numbers: 961231, exp. 2/8/2024; g) THLH71C, UDI/DI 00191072155108, Lot Numbers: 756221, exp. 9/1/2023; 708221, exp. 10/19/2023; h) UIHY16AA, UDI/DI none, Lot Numbers: 931231, exp. 3/10/2024; 896231, exp. 4/14/2024; i) WELP67C, UD |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0828-2024%22