FDA Device recall Z-0828-2024

Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LOWER PELVIC PACK, kit number AGLP52L

On 2024-02-07 the FDA classified a Class II recall of Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LOWER PELVIC PACK, kit number AGLP52L by American Contract Systems, citing during an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting doc….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0828-2024
ClassificationClass II
Statusongoing
Initiated2023-11-29
Published2024-02-07
Last updated2026-09-13 11:57 UTC
CompanyAmerican Contract Systems
DistributionUS

Product

Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LOWER PELVIC PACK, kit number AGLP52L; b) OB/GYN DAVINCI PACK A RG BASIN, kit number HGOB98G; c) DAVINCI GYN PACK, kit number HHDV17; d) KIT, GYNE ROBOTIC, kit number MMGR44D; e) KIT, MAJOR LITHOTOMY, kit number MMJL16E; f) GYN PACK, kit number SMGY46; g) Lithotomy Pack, kit number THLH71C; h) ASC HYSTEROSCOPY PACK (PS 020634), kit number UIHY16AA; i) GYN LAPAROSCOPY HPES, kit number WELP67C; j) GYN LAPAROSCOPY HPES, kit number WELP67C-01; k) GYN LAPAROSCOPY HPES, kit number WELP67C-02

Reason

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Identifiers

code_info
a) AGLP52L, UDI/DI 00191072144119, Lot Numbers: 933231, exp. 3/8/2024; 942231, exp. 2/27/2024; 947231, exp. 2/22/2024;…a) AGLP52L, UDI/DI 00191072144119, Lot Numbers: 933231, exp. 3/8/2024; 942231, exp. 2/27/2024; 947231, exp. 2/22/2024; 966231, exp. 2/3/2024; 980231, exp. 1/20/2024; 681221, exp. 11/15/2023; 707221, exp. 10/20/2023; b) HGOB98G, UDI/DI none, Lot Numbers: 667221, exp. 11/29/2023; 977231, exp. 1/23/2024; 952231, exp. 1/31/2024; 896231, exp. 1/31/2024; c) HHDV17, UDI/DI 00191072170125, Lot Numbers: 766221, exp. 8/22/2023; 731221, exp. 9/26/2023; 682221, exp. 11/14/2023; 638221, exp. 12/28/2023; 946231, exp. 2/23/2024; d) MMGR44D, UDI/DI none, Lot Numbers: 709221, exp. 10/18/2023; 650221, exp. 12/16/2023; 891231, exp. 4/19/2024; e) MMJL16E, UDI/DI none, Lot Numbers: 989231, exp. 1/11/2024; 920231, exp. 3/21/2024; f) SMGY46, UDI/DI 00191072188458, Lot Numbers: 961231, exp. 2/8/2024; g) THLH71C, UDI/DI 00191072155108, Lot Numbers: 756221, exp. 9/1/2023; 708221, exp. 10/19/2023; h) UIHY16AA, UDI/DI none, Lot Numbers: 931231, exp. 3/10/2024; 896231, exp. 4/14/2024; i) WELP67C, UD