FDA Device recall Z-0827-2016

Evita Infinity V500 Ventilator

On 2016-03-02 the FDA classified a Class I recall of Evita Infinity V500 Ventilator by Draeger Medical, citing the battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0827-2016
ClassificationClass I
Statusterminated
Initiated2016-02-04
Published2016-03-02
Last updated2026-09-13 11:54 UTC
CompanyDraeger Medical
DistributionUS

Product

Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.

Reason

The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.

Identifiers

code_infoCatalog #s 8416400 and 8417400

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.