FDA Device recall Z-0824-2015

Bone Screw, self-tapping The Trilogy shells are available in three configurations

On 2014-12-24 the FDA classified a Class II recall of Bone Screw, self-tapping The Trilogy shells are available in three configurations by Zimmer Manufacturing B V, citing analysis of returned complaint devices, product in inventory, and device manufacturing records has identified tool marks isolated to one specific machining mill to be the cause of….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0824-2015
ClassificationClass II
Statusterminated
Initiated2014-11-21
Published2014-12-24
Last updated2026-09-13 11:53 UTC
CompanyZimmer Manufacturing B V
DistributionUS

Product

Bone Screw, self-tapping The Trilogy shells are available in three configurations; uni, multi, and cluster-holed. The multi and cluster-holed shells allow for the use of screws to provide additional fixation and security, particularly in those case where acetabular bone stock is deficient. The Trilogy Acetabular System is indicated for either cemented or uncemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.

Reason

Analysis of returned complaint devices, product in inventory, and device manufacturing records has identified tool marks isolated to one specific machining mill to be the cause of the fractures. .

Identifiers

code_info
Item Number Lot Number 00-6250-065-35 62748089 00-6250-065-35 62754329 00-6250-065-35 62793494 00-6250-065-35 62793…Item Number Lot Number 00-6250-065-35 62748089 00-6250-065-35 62754329 00-6250-065-35 62793494 00-6250-065-35 62793495 00-6250-065-35 62793501 00-6250-065-35 62793502 00-6250-065-35 62793503 00-6250-065-35 62813612 00-6250-065-35 62813613 00-6250-065-25 62784617 00-6250-065-25 62784618 2 00-6250-065-25 62784619 00-6250-065-25 62784621 00-6250-065-25 62784622 00-6250-065-25 62818701 00-6250-065-25 62818702 00-6250-065-25 62818703 00-6250-065-25 62818707 00-6250-065-25 62818709 00-6250-065-25 62818710 00-6250-065-25 62825611 00-6250-065-25 62836984