Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson Centrifugal Pump Model #110192-3

On 2020-01-29 the FDA classified a Class II recall of Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson Centrifugal Pump Model #110192-3 by Isopure, citing there is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during us….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0822-2020
ClassificationClass II
Statusterminated
Initiated2019-04-19
Published2020-01-29
Last updated2026-09-13 11:55 UTC
CompanyIsopure
DistributionAL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, VA

Product

Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson Centrifugal Pump Model #110192-3

Reason

There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.

Identifiers

code_infoIsoMix Pro UDI: (01)00852456007085; ISOMIX Serial Number IMP-838703 with Pump SN 180210D18