FDA Device recall Z-0822-2020
Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson Centrifugal Pump Model #110192-3
- FDA Device
- Class II
- terminated
- Published 2020-01-29
On 2020-01-29 the FDA classified a Class II recall of Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson Centrifugal Pump Model #110192-3 by Isopure, citing there is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during us….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0822-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-04-19 |
| Published | 2020-01-29 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Isopure |
| Distribution | AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, VA |
Product
Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson Centrifugal Pump Model #110192-3
Reason
There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.
Identifiers
| code_info | IsoMix Pro UDI: (01)00852456007085; ISOMIX Serial Number IMP-838703 with Pump SN 180210D18 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.