FDA Device recall Z-0821-2017
Tibial Component Metal Cased Small Long
- FDA Device
- Class II
- terminated
- Published 2016-12-28
On 2016-12-28 the FDA classified a Class II recall of Tibial Component Metal Cased Small Long by Stanmore Implants Worldwide, citing the recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's st….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0821-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-12-07 |
| Published | 2016-12-28 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Stanmore Implants Worldwide |
| Distribution | US |
Product
Tibial Component Metal Cased Small Long; Tibial Component Metal Cased Small Short; Tibial Component Metal Cased Std Long; Tibial Component Metal Cased Std Short The product is a component is a systems that is intended for the replacement if diseased or deficient bone around the knee joint, distal femur, proximal femur or total femur bone
Reason
The recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's stem.
Identifiers
| code_info | A10337, A11525, A7823, A9990, A11289, A11979, A12134, A7186, A8679, A9743, A10344, A10629, A10791, A10348a…A10337, A11525, A7823, A9990, A11289, A11979, A12134, A7186, A8679, A9743, A10344, A10629, A10791, A10348a, A11198, A11557, A11895, A12090, A12129, A12220, A12223, A12229 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0821-2017%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.