Tibial Component Metal Cased Small Long

On 2016-12-28 the FDA classified a Class II recall of Tibial Component Metal Cased Small Long by Stanmore Implants Worldwide, citing the recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's st….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0821-2017
ClassificationClass II
Statusterminated
Initiated2016-12-07
Published2016-12-28
Last updated2026-09-13 11:54 UTC
CompanyStanmore Implants Worldwide
DistributionUS

Product

Tibial Component Metal Cased Small Long; Tibial Component Metal Cased Small Short; Tibial Component Metal Cased Std Long; Tibial Component Metal Cased Std Short The product is a component is a systems that is intended for the replacement if diseased or deficient bone around the knee joint, distal femur, proximal femur or total femur bone

Reason

The recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's stem.

Identifiers

code_info
A10337, A11525, A7823, A9990, A11289, A11979, A12134, A7186, A8679, A9743, A10344, A10629, A10791, A10348a…A10337, A11525, A7823, A9990, A11289, A11979, A12134, A7186, A8679, A9743, A10344, A10629, A10791, A10348a, A11198, A11557, A11895, A12090, A12129, A12220, A12223, A12229