FDA Device recall Z-0821-2014

Puritan Bennett 840 Series Ventilator PB840 Operators Manual Addendum PN 066009A 09/02 The intended use of the oxygen sensor is to provide…

On 2014-01-29 the FDA classified a Class II recall of Puritan Bennett 840 Series Ventilator PB840 Operators Manual Addendum PN 066009A 09/02 The intended use of the oxygen sensor is to provide… by Nellcor Puritan Bennett Dba Covidien, citing covidien is correcting labeling to clarify the operational life of the oxygen sensor (O2 sensor), PN 4-072214-00 used in the Puritan Bennett" 840, and PN, G-062010-00, used in Pur….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0821-2014
ClassificationClass II
Statusterminated
Initiated2013-01-25
Published2014-01-29
Last updated2026-09-13 11:53 UTC
CompanyNellcor Puritan Bennett Dba Covidien
DistributionUS

Product

Puritan Bennett 840 Series Ventilator PB840 Operators Manual Addendum PN 066009A 09/02 The intended use of the oxygen sensor is to provide the oxygen concentration of the delivered gas at the patient wye (inspiratory-expiratory interface). The oxygen sensor does not control the oxygen level or the flow of gases. The intended use of the devices is as critical care ventilators that provide continuous mechanical ventilation for patients.

Reason

Covidien is correcting labeling to clarify the operational life of the oxygen sensor (O2 sensor), PN 4-072214-00 used in the Puritan Bennett" 840, and PN, G-062010-00, used in Puritan Bennett" 740, and760 ventilator systems. The Operators Manual for PB840 states a nominal life of one year and the Addendum incorrectly states 2 years.

Identifiers

code_infoPB840 Operators Manual Addendum: PN 066009A 09/02

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.