RecallRadar

FDA Device recall Z-0817-2016

MEE-1000A Neuromaster, Model No. Accessories: MS-110B (Electric Stimulation), MS-120BK (Current Stimulator Box) JB-116B, JB-132B (Amplifier…

On 2016-02-24 the FDA classified a Class II recall of MEE-1000A Neuromaster, Model No. Accessories: MS-110B (Electric Stimulation), MS-120BK (Current Stimulator Box) JB-116B, JB-132B (Amplifier… by Nihon Kohden America, citing the MEE-1000A Neuromaster had defective printed circuit boards (PCBs). Gas arrestors on these boards which were intended to provide protection from electrostatic discharge failed….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0817-2016
ClassificationClass II
Statusterminated
Initiated2012-01-19
Published2016-02-24
Last updated2026-09-13 11:54 UTC
CompanyNihon Kohden America
DistributionUS

Product

MEE-1000A Neuromaster, Model No. Accessories: MS-110B (Electric Stimulation), MS-120BK (Current Stimulator Box) JB-116B, JB-132B (Amplifiers) Electromyography. (EMG)

Reason

The MEE-1000A Neuromaster had defective printed circuit boards (PCBs). Gas arrestors on these boards which were intended to provide protection from electrostatic discharge failed to prohibit unintentional current flow.

Identifiers

code_info
Serial Numbers: MS-110B: (17 units) 00013-00019, 00023- 00032 MS-120BK: (6 units) 0009- 00014 JB-116B: (31 units), 0…Serial Numbers: MS-110B: (17 units) 00013-00019, 00023- 00032 MS-120BK: (6 units) 0009- 00014 JB-116B: (31 units), 0008, 00012, 00014, 00016, 00018, 00019, 00020, 00021, 00023, 00025, 00026, 00027, 00028, 00029, 00031, 00032, 00034, 00035, 00039, 00043, 00102,00103, 00122-00128, 00134, 00135 JB-132B: (83 units) 00009, 00011, 00012, 00014, 00015, 00016, 00018, 00019, 00020, 00021, 00023, 00024, 00025, 00026, 00027, 00036, 00037, 00038, 00041, 00042, 00043, 00044, 00046, 00047, 00049, 00061, 00062, 00066, 00067, 00068, 00069,00071, 00072, 00074, 00075, 00082, 00084, 0085, 00086, 00090, 00095, 00099, 00120, 00121-00129, 00131-00147, 00149-00155, 00157, 00161, 00164-00168