FDA Device recall Z-0816-2014

ABL90 FLEX Analyzer Radiometer Medical ApS, ¿kandevej 21 Br¿nsh¿j, Denmark A portable, automated analyzer that measures pH, blood gases, el…

On 2014-01-29 the FDA classified a Class II recall of ABL90 FLEX Analyzer Radiometer Medical ApS, ¿kandevej 21 Br¿nsh¿j, Denmark A portable, automated analyzer that measures pH, blood gases, el… by Radiometer America, citing RADIOMETER become aware that if the ABL90 FLEX is tilted, this may potentially cause a negative bias on sodium (Na+) results.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0816-2014
ClassificationClass II
Statusterminated
Initiated2013-10-15
Published2014-01-29
Last updated2026-09-13 11:53 UTC
CompanyRadiometer America
DistributionAL, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV

Product

ABL90 FLEX Analyzer Radiometer Medical ApS, ¿kandevej 21 Br¿nsh¿j, Denmark A portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood.

Reason

RADIOMETER become aware that if the ABL90 FLEX is tilted, this may potentially cause a negative bias on sodium (Na+) results.

Identifiers

code_infoLot R0375N009, Model #393-090, S/N 393-090R0027N001 - 393-090R0379N010