GE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12

On 2025-01-15 the FDA classified a Class II recall of GE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12 by Ge Medical Systems, citing GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0812-2025
ClassificationClass II
Statusongoing
Initiated2024-11-25
Published2025-01-15
Last updated2026-09-13 11:57 UTC
CompanyGe Medical Systems
Distributionn/a

Product

GE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12; 2) 5555000-5; 3) 5555000-6; 4) A0659PB; 5) to be provided

Reason

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.

Identifiers

code_info
1) 5555000-12: UDI/DI none, Serial Numbers: 082407100532; UDI/DI to be provided, Serial Numbers: 082407160512; 2)…1) 5555000-12: UDI/DI none, Serial Numbers: 082407100532; UDI/DI to be provided, Serial Numbers: 082407160512; 2) 5555000-5: UDI/DI 840682115261, Serial Numbers: 1557XR0205, 352867XRA, 478765CDP1, 509962KV220, 601358OPTX220, 620421P3, 813872OPT3, SB2201700099WK; UDI/DI none, Serial Numbers: 479314220, 00028MOB12, 203694SB3, 205820DP1, 208666KDM, 209334OPT1, 210703M3, 240566PX2, 240964OPT1, 248465DP3, 256HHPORT2, 256HHPORT3, 256MHPORT1, 281MSL220, 281MSL2202, 305KEN220A, 306766DMOB2, 310680PORTDR2, 314251NAV7, 325574OPTITBD2, 330489DIGPORT, 333OPTI2, 333OPTI3, 333OPTI4, 352596OPT1, 352LEEDP1, 352VILDP1, 386FHODP3, 401OPTI1, 405636P2, 405636P240, 405752OPT1, 405752OPT2, 405752OPT4, 405DELCITY220, 407303OPTIMAX220, 408259AP6, 417347DPORT1, 417556P1, 417556P2, 417556P3, 419251DP1, 419251DP2, 419383DP1, 434654OPTIMA, 435CVXR220, 480MAYO220A, 501364OPTI1, 501364OPTI2, 509474DR1, 510437OPT1, 510727OPT3, 512SAMCSB, 513870DP1, 520TMCOPT2, 541426XR220, 559447OPT1, 559784OPT2, 561694OPT1

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0812-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.