FDA Device recall Z-0803-2025

Access Intact PTH assay, a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid hor…

On 2025-01-15 the FDA classified a Class II recall of Access Intact PTH assay, a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid hor… by Beckman Coulter, citing beckman Coulter has identified that some Access PTH reagent packs from lot 339071 may be over- or under-filled in well 0. The dispensed material in well 0 could be above or below….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0803-2025
ClassificationClass II
Statusongoing
Initiated2024-11-26
Published2025-01-15
Last updated2026-09-13 11:57 UTC
CompanyBeckman Coulter
DistributionUS

Product

Access Intact PTH assay, a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma using the Access Immunoassay Systems, Catalog Number A16972

Reason

Beckman Coulter has identified that some Access PTH reagent packs from lot 339071 may be over- or under-filled in well 0. The dispensed material in well 0 could be above or below specification, risking sealing issues, contamination, instrument errors, delays, or inaccurate results with a negative bias and increase in imprecision. The issue was confirmed on 10 April 2024 by Beckman Coulter through an internal non-conformance (NC-INT-66585). The well 0 scale on the filling line began malfunctioning on the fill immediately before PTH. The malfunction was not noticed until the majority of PTH was filled.

Identifiers

code_infoUDI: (01)15099590201937(17)241130(11)231201; Lot 339071