FDA Device recall Z-0800-2015
DRX-Revolution Mobile X-Ray System, MODEL DRXR-1, Service Code 1243, Catalog Number 1060177 (China) and 1019397 (all other countries) --- C…
- FDA Device
- Class II
- terminated
- Published 2014-12-24
On 2014-12-24 the FDA classified a Class II recall of DRX-Revolution Mobile X-Ray System, MODEL DRXR-1, Service Code 1243, Catalog Number 1060177 (China) and 1019397 (all other countries) --- C… by Carestream Health, citing while servicing a unit at a customer site, a field engineer discovered a broken bolt from the carriage assembly. Carestream has identified an issue related to the DRX-Revolution M….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0800-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-11-25 |
| Published | 2014-12-24 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Carestream Health |
| Distribution | US |
Product
DRX-Revolution Mobile X-Ray System, MODEL DRXR-1, Service Code 1243, Catalog Number 1060177 (China) and 1019397 (all other countries) --- Carestream Health, Inc. The DRX-Revolution Mobile X-Ray System is a mobile imaging system that incorporates a self contained x-ray generator, image receptor, imaging display and software for acquiring medical diagnostic images outside of a standard x-ray room. The DRX-Revolution Mobile X-Ray System has a motor for driving the system so that the operator does not have to push the system, but can walk behind the system while it is moving from one location to another.
Reason
While servicing a unit at a customer site, a field engineer discovered a broken bolt from the carriage assembly. Carestream has identified an issue related to the DRX-Revolution Mobile X-Ray System in which bolts that secure the x-ray tube assembly/boom to the device mounting block may break. It has been determined that this issue is a result of an improperly assembled part which holds the x-ray
Identifiers
| code_info | Serial Numbers 101 to 764 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Other recalls by Carestream Health
- DRX Revolution Mobile X-Ray System
FDA Device · Class II · 2023-12-13 · Carestream Health - The DRX-Revolution Mobile X-Ray System is a diagnostic mobile x-ray system utilizing digital radiography (DR) technology. the system consis…
FDA Device · Class II · 2023-11-15 · Carestream Health - DRX-Compass/DR-FIT X-ray Systems with Firmware version 2.3.2.0/2.3.2.4. For use in obtaining diagnostic quality radiographic images to aid…
FDA Device · Class II · 2023-03-08 · Carestream Health - Kodak DirectView DR3000/3500 - Product Usage: permanently installed as a diagnostic system intended to generate and control x-rays for exam…
FDA Device · Class II · 2020-01-22 · Carestream Health - OnSight 3D Extremity System- X-Ray, Tomography Computed System Catalog Numbers: 1743566 (w/ Attached Console) 1743574 (w/ Remote Console) T…
FDA Device · Class II · 2019-05-01 · Carestream Health
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0800-2015%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.