FDA Device recall Z-0798-2016

Shuttle Select Slip-Cath Catheter

On 2016-03-02 the FDA classified a Class I recall of Shuttle Select Slip-Cath Catheter by Cook, citing reports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occlud….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0798-2016
ClassificationClass I
Statusterminated
Initiated2015-10-07
Published2016-03-02
Last updated2026-09-13 11:54 UTC
CompanyCook
DistributionUS

Product

Shuttle Select Slip-Cath Catheter. Angiographic catheter. For use in angiographic procedures as a conduit for the delivery of contrast media and can help physicians diagnose occlusion or stenosis. The catheters are designed for percutaneous introduction into the vascular system over an appropriately sized wire guide.

Reason

Reports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result. Recall expanded 10/07/2015 to include additional catheters.

Identifiers

code_info
All lot numbers of the following Product numbers: SCBR4.5-35-135-P-NS-ANG-SHTL, SCBR4.5-35-135-P-NS-ANG-SHTL-JP, S…All lot numbers of the following Product numbers: SCBR4.5-35-135-P-NS-ANG-SHTL, SCBR4.5-35-135-P-NS-ANG-SHTL-JP, SCBR4.5-35-150-P-NS-0-SHTL, SCBR4.5-35-150-P-NS-ANG-SHTL, SCBR4.5-35-75-P-NS-ANGSHTL

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0798-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.