RecallRadar

FDA Device recall Z-0796-2015

PKG, 3MM X 29CM INSERT, RIGHT ANGLE DISSECTOR, LONG JAW, P/N 0250282105 Laparoscopic Manual Instruments are intended for cutting, holding,…

On 2014-12-24 the FDA classified a Class II recall of PKG, 3MM X 29CM INSERT, RIGHT ANGLE DISSECTOR, LONG JAW, P/N 0250282105 Laparoscopic Manual Instruments are intended for cutting, holding,… by Stryker Endoscopy, citing stryker Endoscopy is recalling Laparoscopic manual instruments and accessories because the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 100….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0796-2015
ClassificationClass II
Statusterminated
Initiated2014-11-17
Published2014-12-24
Last updated2026-09-13 11:53 UTC
CompanyStryker Endoscopy
DistributionUS

Product

PKG, 3MM X 29CM INSERT, RIGHT ANGLE DISSECTOR, LONG JAW, P/N 0250282105 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Reason

Stryker Endoscopy is recalling Laparoscopic manual instruments and accessories because the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods.

Identifiers

code_info844265 1041466 1142561 1243266

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0796-2015%22