RecallRadar

FDA Device recall Z-0794-2020

ONE-LINK Needle-free IV Connector, Product Codes: 7N8300, 7N8301, 7N8310, 7N8330K, 7N8332K, 7N8334K, 7N8370K, 7N8371, 7N8375K, 7N8376K, 7N8…

On 2020-01-15 the FDA classified a Class II recall of ONE-LINK Needle-free IV Connector, Product Codes: 7N8300, 7N8301, 7N8310, 7N8330K, 7N8332K, 7N8334K, 7N8370K, 7N8371, 7N8375K, 7N8376K, 7N8… by Baxter Healthcare, citing continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0794-2020
ClassificationClass II
Statusongoing
Initiated2019-12-12
Published2020-01-15
Last updated2026-09-13 11:55 UTC
CompanyBaxter Healthcare
DistributionUS

Product

ONE-LINK Needle-free IV Connector, Product Codes: 7N8300, 7N8301, 7N8310, 7N8330K, 7N8332K, 7N8334K, 7N8370K, 7N8371, 7N8375K, 7N8376K, 7N8377, 7N8378, 7N8390, 7N8391, 7N8399 - Product Usage: intended for single patient use with a vascular access device for the administration of drugs and solutions without needles, thus eliminating the potential for needle-stick injuries during use. This device is an in-line injection site which can be connected to standard male Luer adapters (e.g., syringes or sets) for the continuous or intermittent fluid administration or the withdrawal of fluids.

Reason

Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector.

Identifiers

code_info
PRODUCT CODE UDI 7N8300 85412091839 7N8301 85412475387 7N8301K 85412499635 7N8310 85412476728 7N8330K 85412613499…PRODUCT CODE UDI 7N8300 85412091839 7N8301 85412475387 7N8301K 85412499635 7N8310 85412476728 7N8330K 85412613499 7N8332K 85412599625 7N8334K 85412613482 7N8370K 85412613505 7N8371 85412091822 7N8375K 85412613475 7N8376K 85412599601 7N8377 85412091808 7N8378 85412091792 7N8390 85412475417 7N8391 85412475400 7N8399 85412091785 7N8399K 85412478227 All Lots

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0794-2020%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.