RecallRadar

FDA Device recall Z-0793-2020

Clearlink System Continu-Flo Administration Sets, Product Codes: 2C8519, 2C8537, and 2C8541 - Product Usage:For the administration of fluid…

On 2020-01-15 the FDA classified a Class II recall of Clearlink System Continu-Flo Administration Sets, Product Codes: 2C8519, 2C8537, and 2C8541 - Product Usage:For the administration of fluid… by Baxter Healthcare, citing continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0793-2020
ClassificationClass II
Statusongoing
Initiated2019-12-12
Published2020-01-15
Last updated2026-09-13 11:55 UTC
CompanyBaxter Healthcare
DistributionUS

Product

Clearlink System Continu-Flo Administration Sets, Product Codes: 2C8519, 2C8537, and 2C8541 - Product Usage:For the administration of fluids from a container into the patients vascular system through a vascular access device.

Reason

Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector.

Identifiers

code_infoProduct Codes: 2C8519, UDI 00085412048994; 2C8537, UDI 00085412048970; and 2C8541, UDI 00085412071107 All Lots

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0793-2020%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.