FDA Device recall Z-0790-2017
OSCOR ADELANTE BREEZEWAY 8F C55¿ S61CM D66CM Catalog # AB081040 Is designed to facilitate the introduction of catheters to any of the heart…
- FDA Device
- Class II
- terminated
- Published 2016-12-28
On 2016-12-28 the FDA classified a Class II recall of OSCOR ADELANTE BREEZEWAY 8F C55¿ S61CM D66CM Catalog # AB081040 Is designed to facilitate the introduction of catheters to any of the heart… by Oscor, citing complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0790-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-10-05 |
| Published | 2016-12-28 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Oscor |
| Distribution | CA, FL, MD, MO, NC, NJ, PR, TX |
Product
OSCOR ADELANTE BREEZEWAY 8F C55¿ S61CM D66CM Catalog # AB081040 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
Reason
Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use.
Identifiers
| code_info | Lot #'s: C1-09714 Exp. February 11, 2018, OR-04657 Exp. April 20, 2018, C1-10900 Exp. July 19, 2018, C1-10899 Exp. July…Lot #'s: C1-09714 Exp. February 11, 2018, OR-04657 Exp. April 20, 2018, C1-10900 Exp. July 19, 2018, C1-10899 Exp. July 19, 2018, C1-11048 Exp. August 3, 2018, C1-11729 Exp. October 4, 2018, C1-11729 Exp. October 4, 2018, OR-04729 Exp. October 29, 2018, C1-12233 Exp. December 21, 2018, C1-12234 Exp. December 22, 2018, C1-12234 Exp. December 22, 2018, C1-09062 Exp. June 24, 2017, C1-09062 Exp. June 24, 2017, C1-11047 Exp. August 3, 2018 and C1-12234 Exp. December 22, 2018 2 |
|---|