FDA Device recall Z-0786-2013
Mx8000 IDT Whole Body Computed Tomography X-ray system Distributed by Philips Medical Systems, Highland Heights, OH 44143 The Brilliance Sc…
- FDA Device
- Class II
- terminated
- Published 2013-03-06
On 2013-03-06 the FDA classified a Class II recall of Mx8000 IDT Whole Body Computed Tomography X-ray system Distributed by Philips Medical Systems, Highland Heights, OH 44143 The Brilliance Sc… by Philips Medical Systems Cleveland, citing the low contrast pins in the Image Quality (IQ) phantom deteriorates over time. This prevents the customers from doing the Daily Image Quality Checks as defined in the current Ins….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0786-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-09-18 |
| Published | 2013-03-06 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Philips Medical Systems Cleveland |
| Distribution | US |
Product
Mx8000 IDT Whole Body Computed Tomography X-ray system Distributed by Philips Medical Systems, Highland Heights, OH 44143 The Brilliance Scanner are computed Tomography X-ray systems with multi-slice capability of up to 16 slices,which are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Reason
The low contrast pins in the Image Quality (IQ) phantom deteriorates over time. This prevents the customers from doing the Daily Image Quality Checks as defined in the current Instructions for Use (IfU).
Identifiers
| code_info | 1. Brilliance CT 16 slice (water): System Code #, 728245, Serial #s:2506, 2518, 2555, 2558, 2559, 2561, 2561, 2562, 25…1. Brilliance CT 16 slice (water): System Code #, 728245, Serial #s:2506, 2518, 2555, 2558, 2559, 2561, 2561, 2562, 2564, 2567, 2573, 2576, 2578, 2582, 2583, 2584, 2585, 2586, 2587, 2589, 2590, 2592, 2594, 2596, 2597, 3014, 3051, 3090, 3091, 3105, 3151, 3217, 3235, 3272, 3287, 3303, 3312, 3335, 3338, 3339, 3340, 3341, 3342, 3343, 3344, 3345, 3346, 3347, 3348, 3349, 3350, 3351, 3352, 3353, 3354, 3355, 3356, 3357, 3358, 3359, 3359, 3360, 3361, 3362, 3363, 3364, 3365, 3366, 3367, 3368, 3369, 3370, 3371, 3372, 3373, 3374, 3375, 3376, 3377, 3378, 3379, 3380, 3381, 3382, 3383, 3384, 3385, 3386, 3387, 3388, 3389, 3390, 3391, 3392, 3393, 3394, 3395, 3396, 3397, 3398, 3399, 3400, 3401, 3402, 3403, 3404, 3405, 3406, 3407, 3408, 3409, 3410, 3411, 3412, 3413, 3414, 3415, 3416, 3417, 3418, 3419, 3420, 3421, 3422, 3423, 3424, 3425, 3426, 3427, 3428, 3429, 3430, 3431, 3432, 3433, 3434, 3435, 3435, 3436, 3437, 3438, 3439, 3440, 3441, 3442, 3443, 3444, 3445, 3446, 3447, 3448, 3449, 3450, 3451, 3452, 3 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0786-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.