ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems

On 2014-01-29 the FDA classified a Class II recall of ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems by Philips Medical Systems Cleveland, citing the recalling firm determined that visual inspections of the fork joints behind the detector that are conducted during six month preventive maintenance (PM) may not detect small c….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0785-2014
ClassificationClass II
Statusterminated
Initiated2013-12-04
Published2014-01-29
Last updated2026-09-13 11:53 UTC
CompanyPhilips Medical Systems Cleveland
DistributionUS

Product

ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems. Distributed by Philips Medical Systems, Cleveland, OH

Reason

The recalling firm determined that visual inspections of the fork joints behind the detector that are conducted during six month preventive maintenance (PM) may not detect small cracks or deficiencies in the fork weldment.

Identifiers

code_info
ADAC ARGUS: Model #882060; Serial #: 1023, 1178, 106785, 136560, 1100005, 9051005, 9304030, 9309011, 9501007, 9509005,…ADAC ARGUS: Model #882060; Serial #: 1023, 1178, 106785, 136560, 1100005, 9051005, 9304030, 9309011, 9501007, 9509005, 9510027, 9510028, 9603042, 9603043, 9604044, 9604047, 9605011, 9607060, 9609065, 9610072, 9704081, 9709090, 9710097, 9711103, 9712105, 9712106, 9801107, 9801109, 9802112, 9803001, 9803116, 9804126, 9804129, 9806136, 9807138, 9807422,9810527, 9901156, 9901160, 9901162, 9908167, 9909172, 9910176, 00709B, 0101003A, 01310B, 01525B, 01920B, 01921B, 01922B, 03041B, 03094B, 03338B, 03368B, 2145-30074A, 7718B, A0103015, A0103213, A0105217, A02100230, A02100231, A02110248, A02110249, A02110250, A02110251, A02120254, A02120256, A03010262, A03010264, A03010265, A03010266, A03010270, A03020274, A03020275, A03030282, A03070283, A03120294, B0003196, B0007200, B0309014, ID3935B, M@10639, M@10645, M@10774, M@10777, M@10877, M@11239, M@11568, M@11640, M@11684, M@11685, M@11729, M@11769, M@11813, M@11947, M@12378-1, M@12462, M@12765, M@13725, M@24470, M@4750028851, M@4750028853, M@5348