FDA Device recall Z-0780-2023

VERIFY Steam Test Pack (20 Tests and 5 controls) Item Number: S3065 Intended for use in installation testing and routine monitoring of stea…

On 2023-01-11 the FDA classified a Class II recall of VERIFY Steam Test Pack (20 Tests and 5 controls) Item Number: S3065 Intended for use in installation testing and routine monitoring of stea… by Steris, citing biological indicator (BI) used to promote growth of the G. stearothermophilus and B. atrophaeus spores was inconsistent in promoting growth throughout lot , some BIs within the lo….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0780-2023
ClassificationClass II
Statusongoing
Initiated2022-12-12
Published2023-01-11
Last updated2026-09-13 11:56 UTC
CompanySteris
DistributionUS

Product

VERIFY Steam Test Pack (20 Tests and 5 controls) Item Number: S3065 Intended for use in installation testing and routine monitoring of steam and ethylene oxide sterilization processes

Reason

Biological indicator (BI) used to promote growth of the G. stearothermophilus and B. atrophaeus spores was inconsistent in promoting growth throughout lot , some BIs within the lot performed as intended while others did not. When used to monitor steam and ethylene oxide sterilization cycles, the potential hazard is that the BI will show no growth indicating that the cycle was effective, even if it may not have been.

Identifiers

code_infoUDI: 10724995023840 lot #230613

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.