FDA Device recall Z-0778-2021
Hot AXIOS Stent and Electrocautery - Enhanced Delivery System 20mm x10mm (OUS) UPN: M00553560 - Product Usage: The stent is intended for im…
- FDA Device
- Class II
- terminated
- Published 2021-01-13
On 2021-01-13 the FDA classified a Class II recall of Hot AXIOS Stent and Electrocautery - Enhanced Delivery System 20mm x10mm (OUS) UPN: M00553560 - Product Usage: The stent is intended for im… by Boston Scientific, citing the rotating luer fitting at the distal end of the delivery system handle may become detached from the nose of the delivery during device preparation, stent delivery or use, resul….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0778-2021 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-12-01 |
| Published | 2021-01-13 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Boston Scientific |
| Distribution | US |
Product
Hot AXIOS Stent and Electrocautery - Enhanced Delivery System 20mm x10mm (OUS) UPN: M00553560 - Product Usage: The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution.
Reason
The rotating luer fitting at the distal end of the delivery system handle may become detached from the nose of the delivery during device preparation, stent delivery or use, resulting in delay or procedure or risk of peritonitis developing from leakage of enteric, pseudocyst, or biliary contents into the peritoneal cavity
Identifiers
| code_info | Lot Number: 26167277, 26171099, 26152171, 26171383, 26171380, 26152172 Exp. Date Range: 10/8/2021 10/12/2021 GTIN: 08714729951100 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0778-2021%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.