RecallRadar

FDA Device recall Z-0776-2017

ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened Micro-Introducer Kit, 45CM SS/T NON-ECHO S PG, Item…

On 2016-12-28 the FDA classified a Class III recall of ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened Micro-Introducer Kit, 45CM SS/T NON-ECHO S PG, Item… by Angiodynamics, citing greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the pote….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0776-2017
ClassificationClass III
Statusterminated
Initiated2016-09-28
Published2016-12-28
Last updated2026-09-13 11:54 UTC
CompanyAngiodynamics
DistributionUS

Product

ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened Micro-Introducer Kit, 45CM SS/T NON-ECHO S PG, Item Number H787065970655, Catalog No.REF 06597065, 2) 4 F Stiffened Micro-Introducer Kit, STF 4F M.I. KIT 45CM NT/T NON-ECHO S PG, Item Number H787065970695, Catalog No.REF 06597069 Indicated for percutaneous introduction of guidewire or catheter into the vascular system following a small 21-gauge needle stick

Reason

Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.

Identifiers

code_infoBatch/Lot: 1) 5066410, 2) 5029429, 5043441, 5074176