RecallRadar

FDA Device recall Z-0775-2024

EVair 03 (Jun-air) air compressor, model numbers: 1609000 and 1609002

On 2024-02-07 the FDA classified a Class I recall of EVair 03 (Jun-air) air compressor, model numbers: 1609000 and 1609002 by Datex Ohmeda, citing GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestatio….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0775-2024
ClassificationClass I
Statusongoing
Initiated2023-12-29
Published2024-02-07
Last updated2026-09-13 11:57 UTC
CompanyDatex Ohmeda
Distributionn/a

Product

EVair 03 (Jun-air) air compressor, model numbers: 1609000 and 1609002

Reason

GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40oC (104oF), at the lowest possible flow condition of 2 L/min (worstcase, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen). GE HealthCare is continuing to evaluate the root cause for these elevated levels of formaldehyde. While these test conditions are not representative of typical clinical use conditions,GE HealthCare is taking this action to further reduce the potential for patient exposure.

Identifiers

code_info
UDI/DI ********, Serial Numbers: CBCS01269, CBCT02328, CBCT03050, CBCT02306, CBCT02301, CBCT02303, CBCT02302, CBCT0208…UDI/DI ********, Serial Numbers: CBCS01269, CBCT02328, CBCT03050, CBCT02306, CBCT02301, CBCT02303, CBCT02302, CBCT02088, CBCT02309, CBCT02304, CBCU00717, CBCU00559

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0775-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.